Experimental validation of a finite-difference model for the prediction of transcranial ultrasound fields based on CT images.

Bouchoux G, Bader KB, Korfhagen JJ, Raymond JL, Shivashankar R, Abruzzo TA, Holland CK

The prevalence of stroke worldwide and the paucity of effective therapies have triggered interest in the use of transcranial ultrasound as an adjuvant to thrombolytic therapy. Previous studies have shown that 120 kHz ultrasound enhanced thrombolysis and allowed efficient penetration through the temporal bone. The objective of our study was to develop an accurate finite-difference model of acoustic propagation through the skull based on computed tomography (CT) images. The computational approach, which neglected shear waves, was compared with a simple analytical model including shear waves. Acoustic pressure fields from a two-element annular array (120 and 60 kHz) were acquired in vitro in four human skulls. Simulations were performed using registered CT scans and a source term determined by acoustic holography. Mean errors below 14% were found between simulated pressure fields and corresponding measurements. Intracranial peak pressures were systematically underestimated and reflections from the contralateral bone were overestimated. Determination of the acoustic impedance of the bone from the CT images was the likely source of error. High correlation between predictions and measurements (R(2) = 0.93 and R(2) = 0.88 for transmitted and reflected waves amplitude, respectively) demonstrated that this model is suitable for a quantitative estimation of acoustic fields generated during 40-200 kHz ultrasound-enhanced ischemic stroke treatment.

Keywords:

Parietal Bone

,

Temporal Bone

,

Humans

,

Tomography, X-Ray Computed

,

Ultrasonography

,

Acoustics

,

Models, Biological